SV-10 SV-100 Vibro Viscometer INSTRUCTION MANUAL 1WMPD4000646E
83. DISPLAY AND KEYS 3-1 Display Name Description Standby mode Displays [- - - - -]. Measurement mode Displays the viscosity value in
93-2 Keys Key Description ON:OFF Power Turns the power on and off. When the power is turned on, the viscometer enters the standby mode ( [- -
103-3 Displaying the Viscosity Values The viscosity values are displayed as below, depending on the unit selected and the viscosity range. The corre
11 3-3-2 SV-100 Use the MODE key to switch between Pa⋅s (Pascal second) and P (Poise). The unit selected at the factory before shipment is Pa⋅s. U
124. PRECAUTIONS To get the optimum performance from the viscometer and acquire accurate measurement data, note the following: 4-1 General Precautio
134-2 During Use To level the surface of the sample; adjust the leveling feet so that the center of the narrow part of the right and left sensor pl
144-3 After Use Remove any residual sample material from the sensor plates, temperature sensor and protector using alcohol. Using the sensor plates
154-4 Measuring the Absolute Value of Viscosity The SV Series Sine-wave Vibro Viscometer, as a measuring principle, detects the product of viscosity
164-4-2 At Calibration When calibrating, enter the product of the absolute viscosity value and the density of the standard viscosity fluid used for c
175. MEASUREMENT 5-1 Preparing the Sample 1 Pour the sample into the cup until its surface reaches between the level gauges. The level gauges indic
© 2008 A&D Company Ltd. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated into any l
18NO Note Use the protector in the position as shown on the left below. If the protector is not used with the SV-10, a measurement error may occur,
195-2 Basic Measurement Procedure 1 With the power turned off, press the ON:OFF key. When the display is in the standby mode, press the START ke
205-3 Changing Units The units of viscosity available are: mPa⋅s (millipascal second), Pa⋅s (Pascal second), cP (Centipoise), P (Poise). The cor
216. VISCOSITY CALIBRATION Viscosity calibration is to correct the viscosity value. Two calibration methods are available: one-point calibration
226-2 Calibration Procedure Note As to the correction value used for one-point calibration and two-point calibration, enter the product of the absol
236-2-1 One-point Calibration 1 In the standby mode, press and hold the HOLD key to enter the calibration mode. " Cal " appears. 2 Selec
246-2-2 Two-point Calibration 1 In the standby mode, press and hold the HOLD key to enter the calibration mode. " Cal " appears. 2 Sele
25 9 When the measurement of the first point has completed, clean the sensor plates, temperature sensor and protector and prepare the second standar
266-2-3 Simplified Calibration Using Purified Water (SV-10 only) 1 Place the purified water in the sample cup. 2 Press the START key to measure
27 Reference data: Theoretical viscosity value (Viscosity × Density) of the purified water at various temperatures Temperature (°C) Viscosity × D
1CONTENTS 1. INTRODUCTION ...
287. FUNCTION SETTING The viscometer, by selecting functions to be used in the function setting, can specify the performance appropriate to the usage
29 Example of the function setting procedure Using the SV-10, to change the setting of "Unit upon power-on (Unit)" to the viscosity: cP (Ce
307-2 Details of the Function Items Function item Parameter Description Cladj Date/Time Sets the order of the date (YMD,MDY,DMY) and the date/ti
317-3 Description of Items Date/Time (Cladj) The upper two digits of the year are not displayed. For example, the year 2003 is displayed as "
32 Changing the date The date is changed in the selected displaying order. The following is an example when the order of " y " (Year), &quo
33 Changing the time 12 The current time is displayed. Press the MODE key to display the current date. 13 Press the PRINT key to select the setti
34 Condition (Cond) The stability of the viscosity measurement results can be adjusted, taking ambient conditions such as vibration into consideratio
35 Decimal Point (pnt) Parameter Settings Description 0 • Dot "." 1 Comma "," The decimal point format for the displayed measu
36 Data Output Format (type) The output format appropriate for the device connected to RS-232C can be selected. Parameter Settings Description 0 A
37 Date/time Output (5-td) Parameter Settings Description 0 No output 1 • Output With D.P. format or CSV format selected, whether or not to add the
2 8. CONNECTION TO A PERSONAL COMPUTER...49 9. CONNECTION TO A PRINTER ...
38 Setting the ID number 1 In the standby mode, press and hold the MODE key to enter the function setting mode. " Cladj " appears. 2 Pr
39 Initialization (Clr) Restores the following data to the default setting. Function setting Calibration data After initialization, check the vis
407-4 Data Output Format Examples 7-4-1 A&D Standard Format Used with the printer MODE 1 or MODE 2 when the optional compact printer, AD-8121B
41 SV-100 output format example Viscosity unit Display Output format Remarks L Pa・s OL,-99999999Pas Below measuring range error 1.00 Pa ・s
427-4-2 D.P. Format Used with the printer MODE 3 when the optional compact printer, AD-8121B is connected. With "prt 0" or "prt 1&qu
43 Printing format example (4) Function setting (√=Output Blank=No output) 5-at 1 Measurement elapsed time √ 5-td 0 Date/time Remarks Device ID i
44 Output format example (1) With ID number, date and time added Function setting (√=Output Blank=No output) 5-td 1 Date/time √ 5-ed 3 Device ID n
45 SV-100 output format example Viscosity / Temper-ature Display Output format example Remarks L Pa・s LAB-12,2003/03/19,12:34:56,+025.67,C,+0
46 Output format example (2) With date and time added Function setting (√=Output Blank=No output) 5-td 1 Date/time √ 5-ed 0 Device ID number SV-
477-4-4 RsVisco Format Used with the graphing program RsVisco. Measurement data are output in comma separated format. When a comma is selected as the
31. INTRODUCTION This manual describes how the SV series viscometer works and how to get the most out of it in terms of performance. Read this manu
48 SV-100 output format example Viscosity / Temper-ature Display Output format example Remarks L Pa・s +0000.000, Pa s,+025.67,C Zeroes are o
498. CONNECTION TO A PERSONAL COMPUTER Using the standard RS-232C serial interface and the RS-232C cable provided, the viscometer can be connected t
509. CONNECTION TO A PRINTER The viscometer can be connected to the optional compact printer, AD-8121B using the standard RS-232C interface and th
5110. SERIAL INTERFACE RS-232C Serial Interface Transmission system EIA RS-232C Transmission form Asynchronous, bi-directional, half duplex Data fo
5211. COMMAND LIST The viscometer can be controlled by the following commands from the computer. Add a terminator ( 0Dh、0Ah ) to each command. Comm
5312. TROUBLESHOOTING The viscometer is a precision instrument. When the measuring environment or the measurement method is inadequate, correct value
54How to remove and attach the surface locator plate Removing Loosen the screw and remove the surface locator plate from the temperature sensor. A
55When measurement values are not correct: Has the sample surface been adjusted to the center of the narrow part of the sensor plates? Adjust th
56 When more precise measurement is required: When the viscometer is installed for the first time or is moved to another location, plug in the AC a
5713. ERROR DISPLAY Error display Description Above measuring range error The viscosity value exceeds the upper limit of the viscosity measuring r
5814. SPECIFICATIONS Measurement method SV-10 SV-100 Measurement method Sine-wave Vibro Viscometer using the Tuning-fork Vibration method Vibration
59 *1 When a sample cup of 45 mL is used. *2 Repetitive measurement with the sensor plates remaining in the sample *3 The value after calibratio
6015. OPTIONAL ACCESSORIES List of Optional Accessories (sold separately) Name Number Standard viscosity fluid (JS2.5) *1 AX-SV-31-2.5 Standard
61AX-SV-35 Glass Sample Cup (Approx. 13 mL) • Used when organic solvents will be measured. • The AX-SV-37 consists of : Glass sample cup (Capacity:
62AX-SV-37 Water Jacket Assembly • Used, in combination with a commercially available constant temperature bath for heating medium circulation, to m
63 AX-SV-38 Glass Storage Container • Used to store sample fluids. • The viscosity of the sample stored in the container can be measured as it
6416. EXTERNAL DIMENSIONS Whole View Unit: mm Detailed View of the Sensor Unit Unit: mm
51-2 Features High accuracy The Sine-wave Vibro Viscometer achieves a high measurement accuracy of 1%*1 (repeatability) over the full range. *1 R
62. UNPACKING THE VISCOMETER 2-1 Unpacking The viscometer is a precision instrument. Unpack the viscometer carefully. Keep the packing material t
72-2 Installing the Viscometer Install the viscometer as follows: 1 Connect the display unit to the main unit using the connection cable.