Initialization (Clr)
Restores the following data to the default setting.
Function setting
Calibration data
After initialization, check the viscosity value and perform calibration as necessary. (Refer to
6. VISCOSITY CALIBRATION").
1 In the standby mode, press and hold the
MODE key to enter the function setting
mode. " Cladj " appears.
2 Press the MODE key to select " Clr ".
3 Press the PRINT key to display " Clr no ".
4 Press the START key to select " Clr go ".
5 Press the PRINT key to execute initialization.
After " end ", the next item is displayed.
Initialization has completed.
6 Press the STOP key to return to the standby
mode.