*1 When a sample cup of 45 mL is used.
*2 Repetitive measurement with the sensor plates remaining in the sample
*3 The value after calibration using a standard viscosity fluid at a temperature range
between 20°C and 30°C with no condensation.
In a measurement that takes a long time, perform calibration using a standard viscosity
fluid or purified water periodically, as necessary.
*4 The unit switches to Pas.
*5 The operating temperature of each standard and optional accessory is as follows:
Sample cup (Standard and optional) AX-SV-33: 0 to 120°C
Small sample cup (Optional) AX-SV-34: 0 to 120°C
Glass sample cup (Optional) AX-SV-35: 0 to 230°C
Water jacket assembly (Optional) AX-SV-37: 0 to 100°C
Glass storage container (Optional) AX-SV-38: 0 to 180°C (Lid: 80°C max.)
Take the operating temperature shown above into considerations when using the above
items.